Emerson 5081-P-FF/FI Satellite Radio User Manual


 
26
MODEL 5081-P-FF/FI START-UP - pH
START-UP CONFIGURATION AND CALIBRATION — pH
To perform a two-point buffer calibration, follow these steps:
1. Place the pH sensor in the first buffer solution.
2. On the remote, press CAL, ENTER, ENTER.
3. Use the arrow buttons to select the correct buffer value. Press ENTER.
4. Rinse the sensor and place it in the second buffer solution. Press ENTER.
5. Use the arrow buttons to select the correct buffer value. Press ENTER.
MENU
Sub-menu
PROMPT
Diag Message
FIGURE 14. Menu Tree for pH
PROGRAM
CALIBRATE
GIMP 1000
5081-P-FF
SoFt
InPut 58.9
FAULtS
nonE
rIMP 10
dIAGnOStIC
tEMP
bUFFEr
dISPLAY
ISOPOtntAL
LinE FrEq PAMP=trAnS
CALIbrAtE
Std
tEMP AdJ
tEMP 25.0
tAUtO On
tMAn 25.0
tYPE PH
tEMP C
OUtPUt Cur
COdE 000
bAUtO On
bUFFEr Std
StAbiLiSE
tIME 10
deLtA 00.02
tCOEF 00.00
ISO 07.00
Snr 07.00
LinE 60
dEFAULt
FActOrY
YESNO
rOFFSt 060
dIAG OFF
IMPtC ON
GWH 1000
GFH 1500
GWL 020
GFL 010
CAL 000
rEF LO
rFH 140
rWH 040
rWL 000
rFL 000
CAL bF1
bF 1
bF1 4.01
CAL bF2
bF 2
bF2 10.01
DIAGNOSE
Std 7.00
SLOPE 59.01
SLOPE OFFSt
HArd