Emerson 5081-P-FF/FI Satellite Radio User Manual


 
25
MODEL 5081-P-FF/FI INFRARED REMOTE CONTROLLER
INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONS
The infrared remote controller is used to calibrate and program the transmitter and to display diagnostic mes-
sages. See figure below for a description of the function of the keys.
Hold the IRC within 6 feet of the transmitter, and not more than 15 degrees from the center of the display window.
Infrared Remote Controller.
RESET - Press RESET to end the cur-
rent operation and return to the main
display. Changes will NOT be saved.
RESET does NOT return the transmit-
ter to factory default settings.
CAL - Press CAL to access the calibra-
tion menu.
PROG - Press PROG to access the pro-
gram menu.
DIAG - Press DIAG to read diagnostic
messages.
HOLD - Press HOLD to access the
prompts used for turning on or off the
hold function.
ARROW KEYS - Use
and
keys to
increase or decrease a number or to
scroll through items in a list. Use the
or
keys to move the cursor across a
number. A flashing word or numeral
shows the position of the cursor.
ENTER - Press ENTER to move from a
submenu to the first prompt under the
submenu. Pressing ENTER also stores
changes in memory and advances to the
next prompt.
NEXT - Press NEXT to advance to the
next submenu or to leave a message
screen.
EXIT - Press EXIT to end the current
operation. Changes are NOT saved.
REMOTE CONTROL
INTRINSICALLY SAFE EQUIPMENT
HAZARDOUS AREA LOCATIONS:
CLASS I, DIV 1, GP A, B, C, D
CLASS I, DIV 2, GP A, B, C, D
T3C Tamb = 40°C T3 Tamb = 80°C
1.5Vdc AAA BATTERIES
EVEREADY E92/1212
DURACELL MN2400/PC2400
IS/I/1/A,B,C & D
NI/I/2/A,B,C & D
T4 Tamb = 40°C
T3A Tamb = 80°C
Baseefa02ATEX0198
II 1G EExia IIC T4 1180
1.5Vdc AAA BATTERIES
EVEREADY E92/1212
DURACELL MN2400/PC2400
ROSEMOUNT ANALYTICAL 92606 USA
WARNING:
TO PREVENT IGNITION
CHANGE BATTERIES IN
A NONHAZARDOUS AREA
ONLY
IRC - INFRARED REMOTE CONTROL
LR 34186
Exia
SUBSTITUTION OF
COMPONENTS MAY
IMPAIR INTRINSIC SAFETY
PN 23572-00
YEAR