6-4
Static Pressure Zero Effect
Rosemount 1154DP:
Per 1,000 psi (6.89 MPa):
Rosemount 1154HP:
Per 1,000 psi (6.89 MPa):
Static Pressure Span Effect
Effect is systematic and can be calibrated out for a particular pressure before
installation. Correction uncertainty is ±0.5% of input
reading/1,000 psi (6.89 MPa).
Power Supply Effect
Less than 0.005% of output span/volt
Load Effect
No load effect other than the change in voltage supplied to
the transmitter
Mounting Position Effect
No span effect; zero shift of up to 1.5 inH
2
O (372 Pa) which can be calibrated
out
Response Time
Fixed time constant (63%) at 100 °F (37.8 °C) as follows:
Adjustable damping is available through a special
N option.
FUNCTIONAL
SPECIFICATIONS
Service
Liquid, gas, or vapor
Output
4–20 mA dc
Power Supply
Design limits are as shown in Figure 2-3 on page 2-5. See qualification report
D8400102 for additional detail.
Range Code Static Pressure Zero Effect
4, 5 ±0.2% of upper range limit
6–8 ±0.5% of upper range limit
Range Code Static Pressure Zero Effect
All Ranges ±0.66% of upper range limit
Range Code Response Time
4 0.5 seconds or less
all others 0.2 seconds or less