Rosemount 1154
Drift
±0.2% of upper range limit for 30 months
Range Code 0: ±(0.3% of upper range limit)
Temperature Effect
Range Codes 4–9:
±(0.75% upper range limit +0.5% span) per 100 °F (55.6 °C) ambient
temperature change
Range Code 0:
±(1.13% upper range limit +0.5% span) per 100 °F (55.6 °C) ambient
temperature change
Overpressure Effect
Rosemount 1154DP:
Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:
Rosemount 1154HP:
Maximum zero shift after 3,000 psi (20.68 MPa) overpressure:
Rosemount 1154GP:
Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:
After 4,500 psi (31.0 MPa) overpressure:
After 6,000 psi (41.37 MPa) overpressure:
Range Code Overpressure Effect
4 ±0.25% of upper range limit
5 ±1.0% of upper range limit
6, 7 ±3.0% of upper range limit
8 ±6.0% of upper range limit
Range Code Overpressure Effect
4 ±1.0% of upper range limit
5 ±2.0% of upper range limit
6, 7 ±5.0% of upper range limit
Range Code Overpressure Effect
4 ±0.25% of upper range limit
5–8 ±1.0% of upper range limit
Range Code Overpressure Effect
9 ±0.5% of upper range limit
Range Code Overpressure Effect
0 ±0.25% of upper range limit