Instruction Manual V1.2 - 11/2006
4. Instrument properties
Microtome
Rotary microtome
Section thickness setting 0.5 - 300 μm
Overall specimen feed 25 mm
Vertical stroke 59 mm
Specimen retraction 50 μm
Maximum specimen size 40 x 55 mm
Specimen orientation 8° (x-, y-, z-axis)
Trimming (6 thickness settings) 5 - 150 μm
Motorized coarse feed
- slow 500 μm/s
- fast 1,000 μm/s
Cryocabinet
Dimensions
Width (w/o handwheel) 800 mm
Width (including handwheel) 882 mm
Depth 766 mm
Height (arm rest level) 840 mm
Overall height 1040 mm
Weight
(incl. microtome, w/o specimen cooling) ca. 180 kg
Fluorescent light lamp (cryochamber illumination):
50 Hz version: Osram Dulux S 11 W/21
Light color ‘LUMILUX hellweiß’
(brilliant white)
60 Hz version: Osram Dulux S 13 W/21
Light color ‘LUMILUX hellweiß’
(brilliant white)
according to CECOMAF:
Liquid temperature 45°C
Evaporation temperature -25°C
2
CM3050S-3 CM3050S-6 CM3050S-7
c-UL - -
120 V AC ±10% 208 V AC ±10% 230 V AC ±10%
60 Hz 60 Hz 60 Hz
1800 VA 1800 VA 1800 VA
35 A eff. 30 A eff. 25 A eff.
II I
T15A T1 T12A T1 T10A T1
22 2
II II II
1800 J/s 1800 J/s 1800 J/s
Sectioning motor (Option):
Sectioning speed
- slow min. 0.1 mm/s
max. 100 mm/s
v
max
210 mm/s
- fast min. 0.1 mm/s
max. 170 mm/s
v
max
210 mm/s
*) Refrigerant and compressor oil to be replaced only by trained and authorized service
personnel!
according to IEC-1010, UL 3101
1