Fluke 434 Stereo System User Manual


 
Specifications
Mechanical21
21-15
Mechanical
Item Customer Specification Additional Information
Height x Width x Depth 256 x 169 x 64 mm 10.1 x 6.6 x 2.5 inch
Weight 2.1 kg (4.7 lbs) Including battery pack excluding
current clamps or test leads
Environmental
Item Customer Specification Additional Information
Temperature
Operating within specifications
Operating with reduced
specifications
Non Operating (Storage)
+15 °C ... +35 °C
0 °C ... +50 °C
0 °C ... +40 °C
-20°C ... +60 °C
battery operation only
with power adapter connected
Maximum Relative Humidity
Non Operating (Storage):
Operating:
0 ... 10 °C
10 ... 30 °C
30 ... 40 °C
40 ... 50 °C
No precipitation (non cond.)
No precipitation (non cond.)
95 % ± 5 %
75 % ± 5 %
45 % ± 5 %
Recovery time two hours
No precipitation (no condensing)
battery operation only
Maximum Altitude
Operating
Non Operating
3000 m (10 000 feet)
12 km (40 000 feet)
Above 2000 m derated
overvoltage category
1000V/CATII, 600V/CATIII,
300V/CATIV
Vibration:
Random
Sinusoidal
0.03 g
2
/Hz
3 g
Operating, maximum limits.
MIL-PRF-28800F, class 2,
3.8.4.1&4.5.5.3.1
MIL-PRF-28800F, class 2,
3.8.4.2&4.5.5.3.2
Shock, functional max 30 g MIL-PRF-28800F, class 2,
3.8.5.1&4.5.5.4.1
Bench handling (operating) yes MIL-PRF-28800F, class 2,
3.8.5.3&4.5.5.4.3
Transit drop 1 meter, see Fluke SOP 39.1,
dated Sept. 22, 1992
Drip proof, Dust resistance IP 51 IEC60529 (2001-02)