Emerson Process Management 396P Stereo System User Manual


 
2
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
Measurements and Ranges *pH: 0-14
ORP: -1500 to 1500 mv
Available pH ACCUGLASS Types GPLR hemi or flat glass
Wetted Materials Titanium, Polypropylene, EPDM, glass; platinum (ORP only)
Process Connection 1 in. MNPT front and rear facing threads
Temperature Range 0-100°C (32-212°F)
Pressure Range-Hemi bulb 100-1135 kPa [abs] (0-150 psig)
Pressure Range-Flat bulb 100-790 kPa [abs] (0-100 psig)
Minimum Conductivity 75 μS/cm, nominal; 100 μS/cm
Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial
396PVP: none - must use mating VP cable
Preamplifier Options 396P: Remote or Integral
396PVP: Remote only
Weight/Shipping Weight 0.45 kg/0.9 kg (1 lb/2 lb)
1.2 SPECIFICATIONS
FIGURE 1-1. Cross Section Diagram of the TUpH Patented Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH
glass bulb. This patented sensor technology ensures superior performance while only requiring minimal maintenance.
PERCENT LINEARITY*
pH range GPLR hemi bulb GPLR flat bulb
0 - 2 pH 94%
2 - 12 pH 99% 98%
12 - 13 pH 97% 95%
13 - 14 pH 92%