![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/1/9b/19baf8cb-9a11-4652-82e9-a8eaacde2517/19baf8cb-9a11-4652-82e9-a8eaacde2517-bg3f.png)
Teledyne Advanced Pollution Instrumentation Model 633 Aethalometer
®
Black Carbon Monitor User Manual
07645A DCN6170 61
5.5Manualstabilitytest
Stabilitytestisconductedwithoutflowtodeterminetheperformanceofthelightsource–
detectorpair.IntheOperation/Generaltabselectthedesiredtimebaseandclick“Stability”
buttontostartthetest.Stabilitytestismarkedbystatus1024.DuringthetestBCvalueswill
becalculatedassumingthereis5LPMairflow.Manuallystopthetestwhenthedesiredtest
durationiscomplete.
Interpretation of the results:
-AverageBCvaluesshouldbeclosetozeroiftheAethalometeriswarmedup(was
measuringforafewhoursjustbeforeconductingthetest).
‐PointtopointvariationofBC(PPBC)iscalculatedasanaverageabsolutedifference
betweentheconsecutiveBCmeasurements.Itdependsonthetimebase,testedchannel
andspot.Forexampleat1stimebasePPBC61=approx.400ng/m
3
.
5.5Manualcleanairtest
Cleanairtestisconductedusingthebuilt‐infiltertodeterminetheperformanceof
Aethalometerunderthestandardflowcondition.IntheOperation/Generaltabselectthe
desiredtimebaseandflow,thenclick“Cleanair”buttontostartthetest.Cleanairtestis
markedbystatus2048.Manuallystopthetestwhenthedesiredtestdurationiscomplete.
Interpretation of the results:
‐AverageBCvaluesshouldbeclosetozeroiftheAethalometeriswarmedup(was
measuringforafewhoursjustbeforeconductingthetest).Itispossibletoobservea
shorttransientduetofiltercompression.
‐PointtopointvariationofBC(PPBC)iscalculatedasanaverageabsolutedifference
betweentheconsecutiveBCmeasurements.Itdependsontheselectedtimebase,flow
andtestedchannel.ForexamplePPBC61=approx.500ng/m
3
at5LPMand1s
timebase.