Teledyne Advanced Pollution Instrumentation Model 633 Aethalometer
®
Black Carbon Monitor User Manual
07645A DCN6170 55
5.4Manualflowmetercalibration
First,ensurethataleaktestiscompleted(Section5.1)priortorunningthiscalibration
procedure.Forthemanualcalibrationaflowmeter,datacableandtheflowcalibrationpadare
needed.PleasenoteflowmeasurementsareinMilliliters,andmaximumflowrateshouldbe
nogreaterthan5LPM(5,000milliliters).
Manualflowcalibrationprocedure:
1. STOPmeasurementsinthe
OPERATION>GENERAL
menu.
2. Inthe
OPERATION>ADVANCED
menupressFLOWCAL.
3. SelectMANUALtostartthe
manualflowcalibration.