A SERVICE OF

logo

Teledyne Advanced Pollution Instrumentation Model 633 Aethalometer
®
Black Carbon Monitor User Manual
07645A DCN6170 25
3. INSTRUMENT INSTALLATION
3.1Unpackingthesystem
Taketheinstrumentoutoftheboxandplaceitonanevensurface.Keepthebackofthe
instrumentaccessible.
Remembertoremoveprotectivecapsontheinletandexhaustportsbeforeoperation!
Figure 6.Removeprotectivecapsontherearpaneloftheinstrumentbeforeswitchingiton!
3.2Thesamplingline
The instrument is supplied with several meters of black conductive sample line tubing
designedtominimizeelectrostaticparticlelossesinthesamplingsystem.Keepthesampling
lineasshortaspossible.Donotattempttosubstitutethesamplingtubingwithtubingfroma
differentmaterial.Ifsubstitutingthesuppliedtubingwithtubingfromothermaterial,please
consulttheinstrumentsupplierfordetails.
Wheninstallingthesampleline,trytoavoidsharpbendsorlong horizontalruns with the
sampletubingaseitheroftheseconditionscan promoteparticle lossesinthe sampletube.
Insulate the sample lines inside the instrument shelter and avoid exposing them to direct