Philips M5066A Radio User Manual


 
H-3
PHILIPS MEDICAL SYSTEMS
E
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE
AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE ONSITE
The OnSite is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the OnSite
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and
the OnSite as recommended below, according to the maximum output power of
the communications equipment.
separation distance according to frequency of transmitter (m)
rated maximum output
power of transmitter (W)
80 MHz to
800 MHz
d = 0.6
P
800 MHz to
2.5 GHz
d = 1.15
P
0.01 0.06 0.115
0.1 0.19 0.36
10.6 1.15
10 1.9 3.64
100 6.0 11.5
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2. The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
NOTE 3. An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to
2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if
it is inadvertently brought into patient areas.
NOTE 4. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.