TROUBLESHOOTING WHILE THE ONSITE IS NOT IN USE
(green Ready light is not on)
behavior possible cause recommended action
chirps or
i-button flashes
The battery power is low or the
SMART Pads cartridge needs to be
replaced.
The OnSite may have been turned
off without a pads cartridge
installed, or the installed pads
cartridge may not have its hard
cover in place.
The training pads cartridge has been
left in the OnSite.
The OnSite has been stored outside
the recommended temperature
range.
The OnSite has detected an error
during a self-test or cannot perform
a self-test, or the Shock button is
damaged.
Press the blue i-button. Replace
the battery or pads cartridge if
instructed.
Make sure the pads cartridge is
properly installed with the hard
cover in place. (See Chapter 5,
“Maintaining the HeartStart
OnSite,” for directions on installing
the pads cartridge.)
Remove the training pads cartridge
and replace it with an Adult or
Infant/Child Pads Cartridge.
Remove the battery for five
seconds then reinstall it to start
the battery insertion self-test. If it
fails, insert a new battery to repeat
the test. If it fails again, do not use
the OnSite. If it passes, store the
OnSite within the recommended
temperature range.
Contact Philips for service if
needed.
no chirping and/or
i-button does not flash
The battery is missing or completely
depleted.
The OnSite may have been
physically damaged.
Remove the battery for five seconds
then reinstall it to start the battery
insertion self-test. If it fails, insert a
new battery and repeat the test. If it
fails again, do not use the OnSite.
Contact Philips for service.