January 2015 LIQ_MAN_ABR_5081C-HT/Rev.G
Installation
*#% ###(%)$#
Inspect the shipping container. If it is damaged, contact
the shipper immediately for instructions. Save the box. If
there is no apparent damage, unpack the container. Be
sure all items shown on the packing list are present. If
items are missing, notify us immediately.
'$))#)(%!.
The Model 5081C-HT display can be rotated 90° left or
right. Disengage the cover lock and remove the front
cover. Remove the three screws holding the PCB stack and
gently lift the display board. Do not disengage the ribbon
cable between the display board and the CPU board.
Rotate the display. The black infrared sensor will be at the
top of the display.
#()!!)$#See Figure 7
1. Although the analyzer is suitable for outdoor use, do
n
ot install it in direct sunlight or in areas of extreme
temperatures.
2. Install the analyzer in an area where vibrations and
electromagnetic and radio frequency interference are
minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot
from high voltage conductors. Be sure there is easy
access to the analyzer.
4. The conduit connections on the sides of the 5081
housing should be sealed to prevent moisture from
entering the enclosure.
5. The transmitter must not be mounted with both
conduit openings at the top.
*'"?E>D9>7"?45<)
INCH
MILLIMETER