Preventive Maintenance
Liner and/or insert care
170
Repacking a split insert
Since, for a split insert, its packing material is discarded in cleaning, the
insert must be repacked.
Note: Repacking with small•diameterglass beads is not recommended:
they are usually contaminated with metal filings due to sieving
procedures used. If they must be used, thorough cleaning (chemical and
physical) is required.
Use a fresh, small amount of a conventional coated packing such as
2% OV•1on 100/120 mesh, Chromosorb W•HP.
Packing is held in place between plugs of silanized glass wool (Part
No. 8500•1572).
Glass
Wool
Packing
35 mm
Packing Requirements, Split Insert
Figure 8-8
To ensure compatibility with standard 2•inchsyringe needles, there
should be at least 35 mm between the top of the insert and the top of
the packed section.