24 Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Installation Guide
Operation Verification
Checking the Output Power
With Modulation
11. Preset the signal generator: Press Preset.
12. Select the factory-supplied waveform
SINE_TEST_WFM:
a. Press
Mode > Dual ARB > Select Waveform.
b. Highlight the SINE_TEST_WFM waveform.
c. Press
Select Waveform.
13. Turn the arbitrary waveform player on: Press
the
ARB softkey to highlight On.
14. Set the frequency to the first value listed in
Table 3-6:
Press
Frequency > 125 > MHz.
15. Set the amplitude to 7 dBm:
Press
Amplitude > 7 > dBm.
16. Configure the power meter as follows:
a. On the power meter, press the
Frequency Cal Fac button.
b. Select a power meter channel (if
applicable).
c. Use the arrow keys to enter the frequency
at which to measure the power.
17. Measure the output power.
18. Repeat steps 14 through 17 for the remaining
frequencies listed in Table 3-6, and confirm
that the power level at each point is within
limits.
Tab le 3-6
N5162A/82A Output Power with
Modulation
Frequency
(MHz)
Amplitude
(dBm)
Limits
(dB)
250 7 ±2
323 7 ±2
512 7 ±2
814 7 ±2
1275 7 ±2
2025 7 ±2
2750 7 ±2
3750 7 ±2
5250 7 ±2