Power Quality Analyzer Model 3945-B
19
3.7 AC Current Probe MN93 Probe (3945-B accuracy included)
NOTE: When installing probes, face the arrow on the probe in the direction of
the load.
Nominal Range: 200AAC for f 1kHz
Measurement Range: 2A to 240AACmax(I>200Anotpermanent)
NOTE: Currents <0.5A will be displayed as zero with this probe.
Probe Output Signal: 5mVAC/AAC
Maximum Clamping Diameter:0.8"(20mm)
Safety: EN 61010-2-032, Pollution Degree 2, 300V Cat. IV, 600V Cat. III,
Reference Conditions:
Ambient temperature 73°F(23°C)±5°F(3°C)
Humidity 20 to 75% of RH
Frequency 48 to 65Hz
Distortion factor <1% no DC current
Magneticeldofexternalorigin <40A/m(earth’smagneticeld)
Accuracy
Primarycurrent(AAC) 2 to 10A 10 to 100A 100 to 240A
Accuracy
(%oftheoutputsignal)
3% ± 1ct 2.5% ± 1ct 1% ± 1ct
Phaseshift(°)
Factors affecting accuracy (% of the output signal)
Conditions Range Error
Temperature 14°to122°F(-10°to50°C)
150 ppm/K or
0.15% per 10K
Relative humidity 10 to 90% < 0.2%
Frequency response 40Hz to 10kHz
40Hzto1kHz:<3%
1to10kHz:<12%
Position of the cable in the jaws
<0.5% to 50/60Hz
Adjacent conductor carrying a
60Hz AC current
Conductor in contact with
the sensor
15mA/A
DC current distortion
<20ADC on the
nominal AC current
<5%
Crest factor distortion
3andpeakcurrent=200A 3%
Overload: Frequencyderatingbeyond1kHz:
I max 1000A x
f (in kHz)
1