![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/7/7e/77ea0db0-9271-4090-8c26-242fd2d40cec/77ea0db0-9271-4090-8c26-242fd2d40cec-bg10.png)
14
Testing the Resistance Function
To verify the accuracy of the resistance function, do the following:
1. Connect the Calibrator to VΩHz and COM on the Meter.
2. Turn the rotary switch to Ω.
3. Apply the inputs for step 1-6 in Table 3.
4. Compare the Meter display readings to the display readings in Table 3.
5. If the display reading falls outside of the range shown in Table 3, the Meter does not meet
specification.
Table 3 Resistance Test :
Step Source Reading
1 560.0 553.9 to 566.1
2 5.600k 5.539k to 5.661k
3 56.00k 55.39k to 56.61k
4 560.0k 553.9k to 566.1k
5 5.600M 5.539M to 5.661M
6 36.00M 35.44M to 36.56M
* Lead resistance on the 400Ω range is not included in error.
Testing the Capacitance Function
The Meter measures capacitance by charging the capacitor with a known Direct current,
measuring the resultant voltage, and calculating the capacitance. If the same capacitance is
measured on an impedance bridge, a different reading may result. This variance is likely to be
Greater at higher frequencies.
To verify the accuracy of the capacitance measuring function, do the Following:
1. Apply the Capacitor to the VΩHz and COM inputs on the Meter. For steps 1-4 in Table 4.
2. Turn the rotary switch to
3. Compare the reading on the Meter display to the reading in Table 4.
4. If the display reading falls outside of the range shown in Table 4, the Meter does not meet
specification.
Table 4 Capacitance Test :
Step Source Reading
1 3.6uF 3.3uF to 3.9uF
2 80.0uF 78.3uF to 81.7uF
3 320uF 312uF to 328uF
4 1000uF 979uF to 1021uF