3M
TM
Ranger
TM
Irrigation Fluid Warming System
English
4
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202456B
Introduction
e 3M
TM
Ranger
TM
irrigation uid warming system is designed to warm irrigation uids and deliver these at
ow rates under gravity up to 867 mL/min when the bag is hung 100 cm above the scope. At these ow rates, the
device maintains uid output temperatures ranging from 33°C to 41°C (Note: is is for room temperature uids
only). It takes less than 2 minutes to warm up to the 41°C set point temperature. e alert points on the Model
247 are 48°C and 50°C.
e uid warming disposable set is sterile, latex-free, single-use-only and is designed to be used with the
reusable Model 247 warming unit. For additional information on Ranger disposable sets visit us online at
rangeruidwarming.com. is manual includes operating instructions and unit specications for the Model
247 irrigation uid warming system. For information about using Ranger irrigation uid warming disposable
sets with the Model 247 irrigation uid warming unit, refer to the “Instructions for Use” included with each
disposable component. e Ranger irrigation uid warming system should only be used in healthcare facilities by
trained medical professionals.
Indications for Use
e Ranger irrigation uid warming system is intended to warm irrigation uids.
Definition of Symbols
ON
OFF
Fuse
CAUTION
Type B Applied Part
Voltage, Alternating Current (AC)
An equipotentiality plug (grounded) conductor other than a protective earth conductor or
a neutral conductor, providing a direct connection between electrical equipment and the
potential equalization busbar of the electrical installation. Please consult IEC 6060-1; 2005 for
requirements.
is system is subject to European WEEE Directive 2002/96/EC.
is product contains electrical and electronic components and must not be disposed of
using standard refuse collection. Please consult local directives for disposal of electrical and
electronic equipment.