Escort 3016B CD Player User Manual


 
16 Technicalspecifications
Page 2
16.1.4.Temperaturerecording
CAPTEUR using environment Ranges
PT100
COUPLE J
COUPLE K
COUPLE T
COUPLE S
COUPLE B
COUPLE E
COUPLE N
COUPLEW5
-200 C to 850 C
-210 C to 1200 C
-250 C to 1370 C
-200 C to 400 C
-50 C to 1760 C
200 C to 1820 C
-250 C to 1000 C
-250 C to 1300 C
0 C to 2320 C
20 C to 1000 C
20 C to 2000 C
20 C to 2000 C
20 C to 500 C
50 C to 2000 C
50 C to 2000 C
20 C to 1000 C
20 C to 1000 C
50 C to 2000 C
Theaccuracyofthethermocouplesisindicatedintheappendix
CompensationofthecoldjunctionofthethermocouplesJ,K,T,S,N,E,W5: 1.25 C
16.1.5.Sampling
Resolution: 14bits
Max.samplingperiod:
Memorymode:1s(i.e.1MHz)
Directmode:200 s(i.e.5kHz)
Max.samplingduration: 10min
16.1.6.Bandwidth
Bandwithat-3dB:
RangeBandwidth
>1V100kHz
>50mV50kHz
20mV30kHz
10mV30kHz
5mV20kHz
Internalanalogicalfilters: 10kHz ,1kHz,100Hz,10Hz
slope:20dB/decade
Softwarefilters: 1Hz,0,1Hz,0,01Hz,0,001Hz
Frequencyrecord:
Sensitivity100mVrmsmin.
Minimumdutyfactor10%.
Frequencybetween10Hzand100kHz.
Accuracy0.2%ofthefullscale