MODEL 5081-P-FF/FI START-UP - ORP
START-UP CONFIGURATION AND CALIBRATION — ORP
To perform a one point standardization, follow these steps.
1. Place the ORP sensor in the ORP buffer solution.
2. On the Remote press CAL, ENTER, ENTER
3. Use the arrow buttons to adjust the value. Press ENTER
4. Press RESET
MENU
Sub-menu
PROMPT
Diag Message
FIGURE 15. Menu Tree for ORP
PROGRAM
CALIBRATE
OFFSt 120
5081-P-FF
FAULtS
nonE
dIAGnOStIC
dISPLAY
LinE FrEq
Std
tEMP ADj
tEMP 25.0
tYPE ORP
tEMP C
OUtPUt CUr
COdE 000
LinE 60
rOFFSt 060
dIAG OFF
IMPtC OFF
rEF LO
rFH 140
rWH 040
rWL 000
rFL 000
Std 1000
DIAGNOSE
rIMP 10 SoFt 00.02 HArd 01
PAMP=trAnS
2511
mV