Agilent Technologies N5161A Stereo System User Manual


 
Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 61
Preliminary Basic Operation
Preliminary Modulating the Carrier Signal
3. Enable modulation of the RF output: Press the Mod On/Off key until the LED lights.
If you enable modulation without an active modulation format, the carrier signal does not
modulate until you subsequently turn on a modulation format.
NOTE To turn modulation off, press the Mod On/Off key until the LED turns off.
When the Mod On/Off key is off, the carrier signal is not modulated, even with an active
modulation format.
4. To make the modulated carrier available at the RF output connector, press the RF On/Off key until
the LED lights.
See also: “Using Analog Modulation (Option UNT Only)” on page 75
“Using Pulse Modulation (Option UNU or UNW)” on page 121
“I/Q Modulation” on page 190
Annunciator indicates active AM modulation
AM modulation format on.
A lit LED indicates that any
active modulation format can
modulate the carrier.