Verification and Performance Tests - 2
21
Model Agilent 6633B Report No _______________ Date __________________
Test Description Minimum
Specs.
Results* Maximum
Specs.
Measurement
Uncertainty
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
− 20 mV
Vout − 6 mV
49.955 V
Vout − 21 mV
__________
__________
__________
__________
+ 20 mV
Vout + 6 mV
50.045 V
Vout + 21 mV
1.7 µV
1.7 µV
717.5 µV
717.5 µV
Load Effect
− 4 mV
__________ + 4 mV
35 µV
Source Effect
− 1.0 mV
__________ + 1.0 mV
35 µV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
0 mV
0 mV
__________
__________
+ 3 mV
+ 0.5 mV
872 µV
50 µV
Transient Response
Voltage in 100 µs
0 mV __________ + 50 mV 8 mV
Constant Current Tests
Current Programming and Readback
Low current (0A) Iout
Readback Accuracy @ Iout
High Current (Full Scale) Iout
Readback Accuracy @ Iout
− 1.0 mA
Iout − 0.25 mA
1.998 A
Iout − 4.3 mA
__________
__________
__________
__________
+ 1.0 mA
Iout + 0.25 mA
2.002 A
Iout + 4.3 mA
15.1 µA
15.1 µA
252.5 µA
252.5 µA
Current Sink Readback
Isink − 4.9 mA
__________
Isink + 4.9 mA
252.5 µA
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ − 20 mA
− 2.5 µA
Iout − 22.5 µA
Iout − 22.5 µA
__________
__________
__________
+ 2.5 µA
Iout + 22.5 µA
Iout + 22.5 µA
0.1 µA
1.7 µA
1.7 µA
PARD (Current Ripple and Noise)
RMS
0 mA __________ + 2.0 mA
250 µA
Load Effect
− 1.0 mA
__________ + 1.0 mA
1.6 µA
Source Effect
− 0.25 mA
__________ + 0.25 mA
1.6 µA
* Enter your test results in this column